OD Plate - Verification instrument
Description
The OD Plate is the measurement verification instrument for microplate photometer MPP-96 HiPo. The instrument is designed to verify the accuracy and precision of measurements of the photometer at 6 levels of nominal optical density: 0.3; 0.6; 1.0; 2.0; 3.0; 4.0 OD. The instrument is supplied with the following verification wavelength range: 405–700 nm.
The Instrument is provided in a shockproof container with an USB flash drive containing:
- Copy of measurement results
- User manual
Specifications
Nominal optical density levels |
0.3; 0.6; 1.0; 2.0; 3.0; 4.0 OD (±0.1 OD) |
Available verification wavelengths range |
405, 414, 450, 480, 492, 515, 540, 550, 560, 568, 580, 594, 620, 630, 650, 690, 700 nm |
Overall dimensions (W×D×H) | Weight |
86 x 128 x 12 mm | 0.3 kg |