Fully automated Nucleic Acid purification System
Description
Automated extraction without liquid handling steps
The GenePure PRO 96 is one of a new generation of automated nucleic acid purification systems built around principles of magnetic bead separation. This instrument combines the simplicity of a single-tube extraction procedure with dozens of compatible magnetic bead-based kits to allow for high recovery and purity from nearly any starting sample type. Common sample types include blood, animal tissue, plant tissue, and cultured cells.
The simple programs can process samples simultaneously - and in a matter of minutes. The protocol basics are:
- lyse cells
- bind DNA/RNA to beads
- wash away contaminants
- elute pure DNA or RNA
Each of these steps occurs in 96 deep-well plates with mixing, binding, releasing, and even incubation temperature fully automated. No step during this process requires centrifugation or liquid handling — greatly reducing the need for preventative maintenance and service contracts. The GenePure PRO is made to run for years of uninterrupted sample preparation.
- Max Sample Volume: 20~1'000μL
- Easy to Operate: Self-designed UI operation interface, easy to understand.
- Efficient Extraction: Adopt full magnetic bar and amplitude dynamic adjustment technology, can deal with all kinds of small magnetic bead and reduce the magnetic bead residue.
- UV Sterilization Lamp: urn on manually or automatically.
- Accurate Temp Control: Using the wrapped deep hole heating module, reduce the temperature difference between the tube temperature and the set temperature, and improve the efficiency of lysis and elution.
- Automatic Blocking: Equipped with door protection function, can prevent the pollution and safety problems caused by opening the door by mistake in the experiment.
- Easy to Maintenance: The magnetic bar can be replaced by the user without calibration.
- RUO or CE-IVD
GenePure Pro PROTOCOL OVERVIEW
Each kit comes with a pre-qualified program, so it’s as simple as loading the program, loading the sample & reagents, and pressing “GO”. The GenePure PRO will do the rest. This automated purification technology is well-proven, and for this reason, it’s even compatible with similar reagent kits from other manufacturers.
With a beautiful touch-screen interface, simple operation, and the ability to customize hardware and software to meet demanding requirements of the wide array of users, this system rises to deliver high-purity and refined nucleic acid purification in any lab.
Extract up to 96 samples simultaneously
The GenePure PRO 96 is designed for working with SBS standard 96-well plates that conform to the dimensional requirements SLAS (Society of Laboratory Automation and Screening). It is a common choice for testing labs that require higher throughput levels, but don’t want to sacrifice on the quality of the extracted nucleic acids.
Why GenePure PRO?
- Protocol time under 30 minutes
- Dozens of compatible kits
- Hands-free, error-free purification
- Low maintenance costs
- Proven technology
Specifications
Model |
NPA-96E |
Sample size |
96 pcs/time |
Magnetic rod (fixed) |
96 pcs |
Sample range |
10μL-1’000μL |
Magnetic bead collection efficiency |
>98% |
Most suitable size of magnetic beads |
0.2~1.0μm |
Lysis temperature range |
Room temperature +5°C – 120°C |
Elution temperature range |
+4°C – 120°C |
Module no. |
6 |
Self-detection function |
Yes |
Anti-contamination |
UV lamp / HEPA filter unit |
Device controller |
10″ built-in colour LCD touch screen, windows system |
Barcode reader |
optional |
Memory |
>1000 programs, unlimited with USB flash drive expansion |
Communication port |
USB/LAN |
Dimensions (L x W x H) | weight |
82 x 49 x 57cm | 63kg |
Warranty |
1 year |
Accessories